EasyManua.ls Logo

Gantner e Series - Page 42

Gantner e Series
78 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
e.bloxx A6-2CF
CONFIGURATION
40
HB_EBLOXX-A62CF_E_V19.doc
Gantner Instruments Test & Measurement GmbH
Example: Pressing Force
Column "Sensor"
In this column it is possible to select a sensor principle, to choose the bridge excitation voltage, to set the coarse
measuring range and to do a linearism of the sensor characteristic. The coarse measurement ranges are set by different
gain factors. The ADC converts the selected range with a resolution of 19 bit respectively ±250,000 steps. So a partial
measuring range like 0.5 mV/V can be selected in the column
Format/Adjustment
.
Note:
When changing the set-up in the column
Sensor
the following columns will be set to default.
The following information and settings are possible:
- Sensor principles: Bridge (means strain gauge, inductive, potentiometer), LVDT
- Bridge excitation: ±5 Veff, ±2.5 Veff, ±1 Veff
- Measuring range: low, medium, high (see table 4.1.)
Example: Bridge sensor, excitation 5 V, Range 2.5 mV/V
- Linearization: 8 point with interpolation between the points.
Note:
It is not possible to change the characteristic of the standard sensors in the database.
Use the function
Copy
and edit to change an existing characteristic or select
New
to put a
new sensor in the database.

Table of Contents