Chapter 4. Error Codes and Troubleshooting
86 PanaFlow™ LC User’s Manual
A6:(S2:3)Ch Not
Calibrated
Error occurs when Analog
Input/RTD(S2:3) are not
calibrated
Calibrate the Analog Input/RTD input. If error
persists after calibration, contact BHGE factory
A7:(S2:4)Ch Not
Calibrated
Error occurs when Analog
Input/RTD (S2:4) are not
calibrated
Calibrate the Analog Input/RTD input. If error
persists after calibration, contact BHGE factory
A10:(S2:3)Input
NotConnect!
Analog Input: Error occurs
when (4-20mA) input is not
connected at Channel (S2:3).
RTD Input: Error occurs when
RTD input is not connected or
temp greater than 390 deg C
at Channel (S2:3)
Check connectivity for Analog Input/RTD input
and RTD temperature. Try power cycling the
meter. If error persists after power cycle, contact
BHGE factory
A11:(S2:4)Input
NotConnect!
Analog Input: Error occurs
when (4-20mA) input is not
connected at Channel (S2:4).
RTD Input: Error occurs when
RTD input is not connected or
temp greater than 390 deg C
at Channel (S2:4)
Check connectivity for Analog Input/RTD input
and RTD temperature. Try power cycling the
meter. If error persists after power cycle, contact
BHGE factory
A12:(S2:3)Ch
OverRange Err!
Exceeds input values. For
analog input (S2:3) greater
than 21mA
Ensure analog input current less than 21mA. Try
power cycling the meter. If error persists after
power cycle, contact BHGE factory
A13:(S2:4)Ch
OverRange Err!
Analog input(S2:4) greater
than 21mA
Ensure analog input current less than 21mA. Try
power cycling the meter. If error persists after
power cycle, contact BHGE factory
A24:Aout(S2:1)OutOfRa
nge!
When output from analog
output(S2:1) exceeds 21 mA
or less than 3.6 mA
Check the flow velocity. If velocity is within
limits and error still persists, contact BHGE
factory
A25:Aout(S2:2)OutOfRa
nge!
When output from analog
output(S2:2) exceeds 21 mA
or less than 3.6 mA
Check the flow velocity. If velocity is within
limits and error still persists, contact BHGE
factory
A31:(S2:3)Ch
UnderRange!
Lesser input values. For
analog input(S2:3) between
3.6 mA to 0.25mA
Check input analog current is between 3.6 mA to
21mA. If error persists, contact BHGE factory
A32:(S2:4)Ch
UnderRange!
Lesser input values. For
analog input(S2:4) between
3.6 mA to 0.25mA.
Check input analog current is between 3.6 mA to
21mA. If error persists, contact BHGE factory
Error Code Error Message Description