PanaFlow™ LC User’s Manual 111
Appendix B. Using the Clamping Fixtures
5. From one of the marks on the top of the pipe, measure around the pipe a distance equal to 1/4 of the
pipe circumference, or a distance that will satisfy the orientation found in Step 1. Use the center
punch to make a mark at this point.
Figure 105: GCF Odd Traverse Installation, Step 5
6. From the other mark on the top of the pipe, measure around the pipe in the opposite direction a
distance equal to 1/4 of the pipe circumference, or the same distance used in Step 5. Use the center
punch to make a mark at this point.
Figure 106: GCF Odd Traverse Installation, Step 6
7. Center one of the blocks over one of the center punch marks on the side of the pipe. Align the block so
that the pressure bolt is directly over the punch mark. Secure the block by wrapping the two steel
straps around both the block and the pipe and tightening the straps.
Figure 107: GCF Odd Traverse Installation, Step 7