Appendix B. Using the Clamping Fixtures
116 PanaFlow™ LC User’s Manual
Figure 115: MCF Odd Traverse Installation, Step 3
4. Using a level and center punch, make two marks on the line drawn in step 3. These marks must be
separated by the transducer spacing distance S, as calculated by the flowmeter.
Figure 116: MCF Odd Traverse Installation, Step 4
5. From one of the marks on the top of the pipe, measure around the pipe a distance equal to 1/4 of the
pipe circumference, or a distance that will satisfy the orientation found in Step 1. Use the center
punch to make a mark at this point.
Figure 117: MCF Odd Traverse Installation, Step 5