Appendix G. Wireless HART Communication
138 PanaFlow™ LC User’s Manual
Set IP Address to 192.168.1.10.
Set mask to 255.255.255.0.
Set Gateway to 192.168.1.12 (Note: this is the IP Address on the PC)
6. Verify operation: Note that HART is a slow protocol, Wireless HART is even slower, so be patient waiting
for the Gateway to see the VECTOR.
a. Look at the Network Overview page to see if there are any "live" devices on the network.
b. When visible switch to the Explorer page to see the VECTOR adapter and Wireless HART Data.