Chapter 3. Programming
56 PanaFlow™ LC User’s Manual
Figure 63: Setting up Option I/O Analog Output
3.5.5b 3Option IO (Slot2): Calibrating Analog Output
To calibrate Option I/O Analog Output, use steps as in section “Log-in and Primary Pages” to navigate to the
Input/Output settings page. You may use a multimeter or DCS/SCADA to calibrate the Analog Output.
Regardless of whether multimeter or DCS/SCADA is used, the steps below remain the same. For better
readability, the steps below only indicate multimeter and does not repeat multimeter or DCS/SCADA.
Note: Optional I/O menu uses Slot:Channel convention for clarity. For Example, Analog Output(S2:1) indicates
Analog Output on Slot 2, Channel 1. Optional I/O is installed in the Slot2 of the electronics stack.
1. Connect the Analog Output as in Figure 64.
2. Turn ON the multimeter (if used) and set it to measure Current (mA) DC. Connect test lead from
positive side (Analog Output Channel 1: I/O 1) or (Analog Output Channel 2: I/O 3) of the optional I/O
Analog Output to the positive terminal of multimeter, and the negative lead to the negative
terminal (Analog Output Channel 1: I/O 2) or (Analog Output Channel 2: I/O 4).
3. Highlight [Option I/O (Slot 2)] and press [ENTER].
4. Then highlight [AO-AO-AI-R-1000-3W] and press [ENTER]. Then highlight [Analog Output(S2:1)] or
[Analog Output(S2:2)] and press [ENTER].
5. Then highlight [4-20mA] and press [ENTER].
6. Scroll down and select [Calibrate] option.
Analog Output (S2:1)
Off
Measurement Type
Channel 1
Channel 2
Channel 3
Composite
Measurement Type/Comp
Analog Output (S2:1)
Base Value
Full Value
Calibrate
Measurement Type
Velocity
Volumetric (Act)
Soundspeed
Volumetric (Std)
Avg Vol Flow Rate
Mass Flow
Mass Flow
Full Value
UNIT: ft/s
3.0000
Calibrate 4mA
Full Value
Full Value
3.0000 ft/s
Analog Output (S2:1)
Calibrate
Base Value
Error Handling
Analog Output (S2:1)
Calibrate
Base Value
Error Handling
Full Value
Calibrate
Error Handling Value
Analog Output (S2:1)
Base Value
Full Value
Error Handling
Low
High
Hold
Other
Measurement Type/Comp
Measurement Type/CH #1
Reynolds#
Gain
Avg Vol Flow Rate
Soundspeed
Measurement Type/CH #1
Std Deviation Gain
Measurement Type
Composite
Channel 1
Channel 2
Channel 3
Velocity
Volumetric (Act)
Mass Flow
Calibrate 4mA
4.000
UNIT: mA
Calibrate
4.00 mA
Value
UNIT: mA
4.00
4-20mA
Full Value
Error Handling
Value
9
2
9
2
9
2
9
2
9
2
9
2
9
2
or
9
2
9
or
9
2
9
or
9
2
9
or
9
2
9
or
9
2
9
Input/Output
Option I/O (Slot 2)
Analog Output (Std)
Digital Output (Std)
Option I/O (Slot 2)
AO-AOAI-R-1000-3W
9
2
Option Comm (Slot 1)
9
2
AO-AOAI-R-1000-3W
Analog Output (S2:1)
RTD Input (S2:4)
Analog Output (S2:2)
Analog Input (S2:3)
Volumetric (Std)
SNR Up
Gain
Std Deviation Gain
Measurement Type/CH #1
SNR Dn
Calibrate 20mA
Percentage of Scale
Operator Access or Higher
[ ]
2
9
[ ]
UNDO
MOVE
[ ]
[ ]
SAVE
ALTER
[ ]
2
9
[ ]
UNDO
MOVE
[ ]
[ ]
SAVE
ALTER
[ ]
2
9
[ ]
UNDO
MOVE
[ ]
[ ]
SAVE
ALTER