5.5 Setting the analysis temperature
The SPR signal is sensitive to changes in temperature. It is important that a constant
temperature is maintained at the sensor chip surface throughout the run. The detection
area housing the sensor chip is maintained at a precisely controlled temperature.
The analysis temperature is fixed at 25°C in the standard instrument. With the optional
Biacore X100 Plus Package, the analysis temperature can be set to 4°C to 40°C (maximum
10°C below ambient temperature).
The analysis temperature is displayed in the status bar at the bottom of the screen.
Unstable temperature is indicated by
a flashing Temperature indicator on the instrument front panel,
a flashing temperature in the Control Software status bar and
a note in the event log in the Control Software during a run.
56 Biacore X100 Operating Instructions 28961142 AD
5 Operation
5.5 Setting the analysis temperature