5.5 Setting the analysis temperature
The SPR signal is sensitive to changes in temperature. It is important that a constant
temperature is maintained at the sensor chip surface throughout the run. The detection
area housing the sensor chip is maintained at a precisely controlled temperature.
The analysis temperature is fixed at 25°C in the standard instrument. With the optional
Biacore X100 Plus Package, the analysis temperature can be set to 4°C to 40°C (maximum
10°C below ambient temperature).
The analysis temperature is displayed in the status bar at the bottom of the screen.
Unstable temperature is indicated by
a flashing Temperature indicator on the instrument front panel,
a flashing temperature in the Control Software status bar and
a note in the event log in the Control Software during a run.
56 Biacore X100 Operating Instructions 28961142 AD
5 Operation
5.5 Setting the analysis temperature

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Biacore X100 and is the answer not in the manual?

GE Biacore X100 Specifications

General IconGeneral
TypeSurface Plasmon Resonance (SPR) System
ApplicationBiomolecular interaction analysis
Detection MethodSurface Plasmon Resonance (SPR)
Temperature Control4-40°C
SoftwareBiacore X100 Control Software
Power Requirements100-240 V, 50/60 Hz
Sample TypeProteins, nucleic acids, small molecules
Flow Rate Range1-100 μL/min
Data AnalysisAffinity, kinetics, concentration, specificity
Flow SystemMicrofluidic
Dimensions60 cm (W) x 50 cm (D) x 45 cm (H)
ThroughputHigh throughput
Sample CapacityUp to 96 samples

Related product manuals