8.1 Specifications
General
ValueParameter
Maximum 15 samples, 24 h unattended operationAutomation
Down to 100 Da in various sample environmentsMolecular weight detection
Injection volume + 20 to 30 µL (application depen-
dent)
Required sample volume
15 sample/reagent vials, volume 1.5 mL
1 water vial, volume 4 mL
Sample/reagent capacity
Typically 5 to 15 minAnalysis time per sample
Typically 0.1 nM for >10 000 Da analytes.
Typically 1 nM for <10 000 Da analytes.
Detection limit
Typical range for kinetics:
10
3
to 10
7
M
-1
s
-1
(for protein-protein interactions)Association rate constant (k
a
):
10
-5
to 0.1 s
-1
Dissociation rate constant (k
d
):
2Number of flow cells
Approximately 0.05 mmFlow cell height
Approximately 0.06 µLFlow cell volume
Instrument
ValueParameter
600 × 530 × 575 mmDimensions w × d × h
(see also Space requirements, on
page 40)
47 kgNet weight
Autorange 100 to 240 V~,
protective earthing
Supply voltage
± 10% from the nominal voltageMaximum voltage fluctuation
50/60 HzFrequency
84 Biacore X100 Operating Instructions 28961142 AD
8 Reference information
8.1 Specifications