6.5 Replacing the mains fuses
WARNING
Disconnect power. Always disconnect power from the instrument
before replacing fuses.
WARNING
For continued protection from fire hazard, replace only with same
type and rating of fuse.
CAUTION
Do not replace the mains fuses if you suspect that there may be a
malfunction in the instrument. Contact your GE service representa-
tive for advice.
Follow the steps below to replace the mains fuses.
ActionStep
Turn off the power to the Biacore X100 instrument.1
Disconnect the mains cord from the mains power inlet.2
Insert a small screwdriver under the tab on the fuse drawer cover.3
Biacore X100 Operating Instructions 28961142 AD 71
6 Maintenance
6.5 Replacing the mains fuses