Clean the needle and wash
station
WARNING
Liquids in the buffer bottles or tubing may be toxic or flammable
or may cause chemical burns or irritation to skin and eyes. Take
appropriate precautions in the event of bottle breakage, accidental
spillage and insecure fitting of tubings to bottles.
WARNING
The injection needle is sharp. Take care when working in the sample
compartment.
Clean the needle and wash station if you see that salt or other deposits have accumu-
lated. You will need deionized water, lint-free tissues, and an hex key to remove the shield
(supplied with the instrument).
Run Tools:More Tools:Clean needle and needle wash station and follow the instructions
on the screen.
Clean the instrument cover
If necessary, clean the cover of the instrument with a moist cloth. Use water or a mild
detergent. The buffer tray and the waste tray can be removed for cleaning. If salt residues
or proteins have accumulated on the buffer bottle cap, clean the cap in deionized water.
Biacore X100 Operating Instructions 28961142 AD 67
6 Maintenance
6.3 Cleaning