K114 Issue No. 9
6 - 11
Druck ADTS 405 User Manual
6
7. In manual mode, any of the normal key-pad functions may be used. When the manual
testing is complete, press EXECUTE TEST PROGRAM and the test program continues
from where it left off. This feature may be disabled by including the KEYLOCK
command in the test program. While in manual mode the display prompts “TPM”.
8. When a "go" command is executed, “GO” flashes in the lower right hand corner of the
screen. At this point the user should record any instrument readings or displayed leak
rates, as required, before continuing by pressing EXECUTE TEST PROGRAM. F4 may be
used to continue at a “go” command as an alternative to pressing EXECUTE TEST
PROGRAM.
Deleting files
If a test program within an ADTS 405 is no-longer required, then it may be deleted.
To delete a test program, proceed as follows:
1. Enter configuration (CONFIG) by pressing and holding F1 and pressing SETUP.
2. At the first menu level of configuration, press the EXECUTE TEST PROGRAM key, then
F2. The display then shows the file name and title of the first file in the list. Use [NEXT]
and [PREV] to step through the files until the display shows the file to be deleted.
3. Select [DEL] to delete the file, then press CLEAR/QUIT to quit from configuration.
4. Select [YES] to confirm quit.
Note: If the deleted program is marked as “AUTO RUN” then an additional screen asks to
confirm if the auto run status should be removed before deletion.
Auto Running a Program
To run a program each time the ADTS 405 is switched on, the “Auto run” feature should be
enabled. The enabled program starts, from the first line, after successful completion of the
ADTS power-up sequence. To set-up the auto run feature, proceed as follows:
1. Enter configuration (CONFIG) by pressing and holding F1 and pressing SETUP.
2. At the first menu level of configuration, press the EXECUTE TEST PROGRAM key, then
F1. The display shows the name of the current file to “Auto run” or “*NONE*” if auto
run is not selected. To change this selection use the [NEXT] and [PREV] keys and when
the correct file is displayed press [SEL]. To de-select auto run press [NONE].