K114 Issue No. 9
Druck ADTS 405 User Manual
5 - 5
5
CAUTION:ONLY SELECT ITEMS DETAILED IN THE FOLLOWING MENUS, INCORRECT SELECTIONS CAN CAUSE
CHANGES
TO SAFETY LIMITS AND THE CALIBRATION DATA OF THE SYSTEM.
4. The system enters the maintenance mode with the following screen display:
(MAINTENANCE)
F1 Data Block
F2 Supervisor EEPROM
F3 More
5. Select F3. The display shows:
(MAINTENANCE)
F1 System
F2 Hardware
F3 Error log
6. Select F3. The display shows:
(MAINTENANCE)
F1 Vent System
F2 Set System Ready
F3 View Temperature
7. Select F1 and when complete, press CLEAR/QUIT three times to reset the
system.
5.5 Fault Diagnosis
If the display shows a warning or error message, refer to the above sections.
If a fault occurs, refer to the fault location chart, Figure 5-1 and Table 5-1 for possible
fault causes and recommended action.