K114 Issue No. 9
6 - 29Druck ADTS 405 User Manual
6
CONFIG, [UNITS]
This function allows the default units to be configured.
The default [AERONAUTICAL] units are used at power-up for the altitude (and airspeed)
display. The default units are also used for the limits entry in CONFIG and entry of any
altitude or airspeed related functions in SETUP or CONFIG e.g., airfield altitude or
altitude correction.
The default [PRESSURE] units are used for limits entry in CONFIG and entry of any
pressure values in SETUP or CONFIG.
The default [TEMPERATURE] units are used for entry of temperature values in SETUP or
CONFIG.
The default [AERONAUTICAL] and [PRESSURE] are set in MINIMUM SETUP, [UNITS].
CONFIG, [LIMITS], [EDIT LIMITS], [EDIT EXISTING]
This function allows existing sets of limits to be configured.
Use [NEXT] or [PREV] to select the set of limits by aircraft name or component/system
name/number when the display shows the required limits, press [SEL] to edit the limit
values.
Use [NEXT] or [PREV] to step through each limit or associated parameter in turn. Enter
a new numeric value to change a limit.
Press CLEAR/QUIT when all required changes have been made and select [YES] to save
changes. The parameters are:
Name Max Ps
Max Alt Min Ps
Min Alt Max Qc
Max CAS Min Qc
Min CAS Max Rate Ps
Max Mach Max Rate Qc
Max ROC ARINC Limits
Max Rate CAS Altitude Correction
NOTES
1 The ADTS 405, when delivered, contains "Standard" and "Civil" limits; to set and
store "Max" limits, for the first time, enter configuration. The "Max" limits depend
on the range of the pitot channel either 850 or 1000 knots, see section 1.3,
operating limits.
2 The system's defined "STANDARD", "CIVIL" and "MAX" limit sets cannot be
redefined. A defined set can be edited but must be saved using a different name.
This method could save time.