xiii
K114 Issue No. 9
ATEX Certified ADTS Hand Terminal
CONDITIONS OF USE
The ATEX certified ADTS hand terminal supplied with the ADTS 405 can be used in zone 2
hazardous areas in accordance with the ATEX certification document and schedule.
ATEX Certificate of Conformity
No. Baseefa05ATEX0192
BASEEFA being an Approved Certification Body, in accordance with Article 14 of the Council
Directive of the European Communities of 18th December, 1975 (76/117/EEC) certifies that
the apparatus has been found to comply with harmonised European Standards:
EN 60079-15: 2003
and has successfully met the examination and test requirements recorded in confidential
report number:
05(C)0154 (Baseefa) dated 12th October 2005
NOTE: Refer to pages 2/2 of the Certificate of conformity for electrical connection parameters.
Rated Voltage = 32Vdc.
Marking detail:
ADTS Hand Terminal
Baseefa05ATEX0192
II 3G IP54
Baseefa05ATEX0192 (EC type examination certificate number)
EEx nA nL IIC T4 (-25°C
<T
amb
<=55°C)
Druck, LE6 0FH, UK
6 - 32V
Power
3W Max
SPECIAL CONDITION OF USE
The power supplies must be isolated when connecting the ADTS hand terminal in the
hazardous area.
The ADTS hand terminal must not be disconnected when energized in the hazardous
area.
The ADTS hand terminal is a non-serviceable component. If the ADTS hand terminal
becomes unserviceable it can only be replaced by another ATEX compliant hand
terminal.