K114 Issue No. 9
Druck ADTS 405 User Manual 3 - 11
3
The ADTS 405 system may now be used but for full specification accuracy and
stability, wait the "WARMUP" period of 15 minutes. The display shows “WARMUP” in the
lower right hand corner, this message clears automatically after the time period.
3.7 Control or Measure Parameter
To change the displayed parameter:
Value parameters - Press the parameter key
e.g., press SPEED/Qc to display airspeed.
Rate parameter - Press the associated parameter key followed by the rate
key for that channel. e.g., display airspeed rate, press
SPEED/Qc then RATE. ROC/Ps Rate may be directly
pressed without first pressing ALT/Ps.
Note: The displayed parameters depend on the last keys pressed.
To display two parameters:
Press each value parameter in turn.
e.g., display altitude and airspeed together, press ALT/Ps then SPEED/Qc.
In control mode, an arrow, at the right-hand end of the aim value, indicates
the last parameter selected. This can be changed by entering a new value.
To display a value parameter together with its rate:
Press the parameter key followed by the associated rate key.
e.g., display airspeed and rate of speed together, press SPEED/Qc then RATE.
See Rate set-up for display details.
Note: If ROC/Ps Rate is pressed, the display automatically shows ALT and ROC.
Aim
A new aim can be entered using the numeric keys. Each digit is displayed as it is
pressed. The existing aim is replaced when the first digit of the new aim is pressed. If
an error is made during the entry of data, press CLEAR/QUIT to restore the original
aim.
Press ENTER to action the new aim.
Note 1: The 000 key can be used as a quick way of entering thousands.
Note 2: The ADTS 405 must be in control mode to enter a new aim (current aim
displayed). If the aim field shows "Leak Measure" press LEAK MEASURE/
CONTROL to enter control mode.