K114 Issue No. 9
6 - 15
Druck ADTS 405 User Manual
6
6.3 SETUP
Pressing the SETUP key causes the system to enter a set-up mode preset in the
configuration mode. The following section describes the three functions of setup: Full,
Minimum and Off. The full set-up allows access to all set-up functions. Minimum set-
up allows access to some set-up functions, the off function prevents access to the set-
up functions.
FULL SET-UP
Full set-up allows access to secondary functions that do not have an assigned key.
Any parameter changed under set-up returns to the default settings on power-down.
The CONFIGuration mode may be used to change the default settings so that the ADTS 405
powers up in the required state. Most set-up parameters are also available under CONFIG.
Additional parameters are only available under CONFIG.
Certain parameters in set-up can be locked in CONFIGuration mode preventing
inadvertent changes such as aircraft limits at power-up with a particular set of limits
that must not be changed. Attempting to change a locked parameter causes the
display to show a warning message.
Pressing the SETUP key makes the set-up functions available. Pressing the function
keys selects the set-up menus, using the appropriate function key selects the required
menu choice. To set-up a parameter key, press SETUP then press the required
parameter key. Each parameter key set-up provides settings to the normal user mode
function of the key. Further menus are used under certain direct key set-up functions.
Note:
The Quick Reference Section of this manual provides details of the set-up menus and keys.
Once a setting has been changed in set-up, use CLEAR/QUIT to return to the preceding
menu. After all required parameters have been changed, repeatedly press CLEAR/
QUIT until the display shows the main pressure display.
The following section describes each set-up function. Intermediate menus are not
described.