K114 Issue No. 9
Reference and Specification
6 - 16
SETUP, [UNITS]
Use [NEXT] or [PREV] to step through the available units until the display shows the
required units. Select [SAVE] and then press [QUIT] to make the selection.
The available units are:
ft and kts (includes Mach)
ft and mph (includes Mach)
m and km/h (m/min) (includes Mach)
m and km/h (m/s) (includes Mach)
m and km/h (hm/min) (includes Mach)
mbar
inHg
mmHg
SETUP, [LIMITS]
The ADTS 405 is preprogrammed with the manufacturer's STANDARD, CIVIL and MAX*
limit sets. Different sets of additional aircraft limits can be programmed using the
CONFIG mode. SETUP, [LIMITS] allows the selection of the set of limits to be used, each
set of limits is identified by its aircraft name.
Note: *The ADTS 405, when delivered, contains "Standard" and "Civil" limits; to set and
store "Max" limits, for the first time, enter configuration. The "Max" limits depend
on the range of the pitot channel either 850 or 1000 knots, see section 1.3,
operating limits.
Use [NEXT] or [PREV] to step through the aircraft names select [SEL] and press QUIT
when the display shows the required name.
Use CONFIG, [LIMITS] to program the sets of limits. Each set of limits includes:
Min Ps
Max Qc
Min Qc
Max Rate Ps
Max Rate Qc
ARINC limits on or off
Altitude correction value
Max ALT
Min ALT
Max CAS
Min CAS
Max Mach
Max ROC
Max Rate CAS
Max Ps
The set of limits used at power-up (default) can be selected using CONFIG, [LIMITS],
[DEFAULT AIRCRAFT]. The default aircraft settings can be locked using CONFIG, [LIMITS],
[LOCK AIRCRAFT] so that the operator cannot change the limits.
inH
2
O 4°C
inH
2
O 20°C
inH
2
O 60°F
psi
hPa
kPa
kg/cm
2
% full-scale