5.7 Testing an Option Facility
The following tests should only be carried out if the serviceability of an option facility is
suspected or when an error occurs.
Testing the IEEE 488 Facility
This IEEE 488 facility requires specialist knowledge of both IEEE 488 communications
and test programming for specified, authorized test procedures.
Only qualified personnel should use this facility, the test procedure must be authorized
by the appropriate engineering authority.
To enable and disable this IEEE 488 facility access the configuration menu. When
enabled, this facility will be available from power-up of the ADTS 405 with the settings
held in non-volatile memory.
Configuring and Enabling the IEEE 488 Facility
To enable the IEEE 488, set the ADTS 405 to the SCPI IEEE option and set to Address 1
as follows:
Press and hold down F1 then press SETUP to enter the configuration menu.
Release both keys.
If requested, enter PIN details.
Press [MORE], [DISPLAY/OPTIONS], [OPTIONS], [IEEE-488], [SCPI/OPT2].
Check the display shows that SCPI is selected, if the settings are correct, press
CLEAR/QUIT until the display shows the main pressure display, run the test
program below.
If the display does not show that SCPI is selected, press [SCPI].
Press CLEAR/QUIT, [DEVICE ADDRESS].
Check the address displayed is 1.
If not, press 1 on the numeric key-pad and then ENTER.
If any changes have been made, press CLEAR/QUIT until requested to accept
the changes, then press ENTER.
The ADTS 405 automatically restarts, when the display shows the main pressure
display, carry out the following test program.