K114 Issue No. 9
6 - 25Druck ADTS 405 User Manual
6
SETUP, HELP
This function provides help on the set-up key.
SETUP, or (nudge keys)
This function allows the increment value of the nudge key to be adjusted for every
parameter.
When SETUP, or are pressed, the display shows the value of nudge for the
parameter last displayed.
Enter a new value to replace the present value.
Use [NEXT] or [PREV] to display other parameters.
Press CLEAR/QUIT when the nudge values of all the required parameters have been
changed.
In aeronautical units, the parameters available are:
ALT CAS/TAS MACH ROC RATE CAS
In pressure units, the parameters available are:
Ps Pt Qc EPR
RATE Ps - RATE Qc RATE EPR
The nudge value can be used for large increments e.g., 100 kts or 5000 ft to "step
through" the cardinal points of a test.
CAUTION:MAKE SURE THE NUDGE INCREMENT IS SET TO A SUITABLE VALUE BEFORE USING THE NUDGE
FACILITY
TO CHANGE THE AIM VALUE.