K114 Issue No. 9
5 - 16
Testing and Fault Finding
Testing the ARINC 429 Option
To enable and disable this ARINC facility access the configuration menu. When
enabled, this facility will be available from power-up of the ADTS 405 with the settings
held in non-volatile memory.
Configuring and enabling the ARINC 429 Option
To enable the ARINC option as follows:
Press and hold down F1 and press SETUP to enter the configuration menu.
Release both keys.
Press [MORE], [DISPLAY/OPTIONS], [OPTIONS], [ARINC-429].
Check the display shows ARINC 429 is selected.
noitcnuF niP ruoloC
ev+tuptuo51wolley
ev-tuptuo61eulb
ev+tupni81neerg
ev-tupni91der
htrae/dnuorGllehs-
Note: If the ARINC 429 option is not fitted the value is replaced by "absent".
If the above procedure is unsuccessful, check the continuity of the ARINC 429 cable.
If faulty, replace the cable.
If the continuity of the cable is correct, return the ADTS 405 to the repair depot.
Alt XXXXX ft
Leak Measure
Bus XXXXX ft
ARINC 429 WARMUP
ADTS 405 mode
value
units of measurement