Performing an Exam
2-28
LOGIQ E9
User Guide
Direction 5454884-100 English
Rev. 1
Slope (Velocity, Time Interval and Acceleration)
To measure two velocity values, the time interval (ms), and
acceleration (m/s
2
):
1. Press Measure. Press the top Trackball key to select Slope;
an active caliper with vertical and horizontal dotted lines
displays.
2. To position the caliper at the start point, move the Trackball.
3. To fix the start point, press Set. The system fixes the first
caliper and displays a second active caliper.
4. To position the second caliper at the end point, move the
Trackball.
5. To complete the measurement, press Set. The system
displays the two peak end point velocities, the time interval,
and the acceleration in the Results Window.
Time interval
To measure a horizontal time interval:
1. Press Measure. Press the top Trackball key to select Time;
an active caliper with vertical and horizontal dotted lines
displays.
2. To position the active caliper at the start point, move the
Trackball.
3. To fix the start point, press Set. The system fixes the first
caliper and displays a second active caliper.
4. To position the second caliper at the end point, move the
Trackball.
5. To complete the measurement, press Set. The system
displays the time interval between the two calipers in the
Results Window.

Table of Contents

Other manuals for GE LOGIQ E9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E9 and is the answer not in the manual?

GE LOGIQ E9 Specifications

General IconGeneral
TypeUltrasound System
3D/4D ImagingYes
Transducer Ports4 active ports
ConnectivityDICOM, USB, Ethernet
ApplicationsAbdominal, Vascular, Musculoskeletal, Small Parts, Urology, Pediatric, Cardiac, Obstetrics/Gynecology
TransducersLinear, Convex, Phased Array, Volume
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
ElastographyYes
Advanced FeaturesB-Flow, Volume Navigation
PortabilityCart-based system
Display19-inch high-resolution LCD monitor

Related product manuals