Performing an Exam
2-26
LOGIQ E9
User Guide
Direction 5454884-100 English
Rev. 1
Doppler Mode Measurements
Four basic measurements can be made in Doppler Mode.
• Velocity
TAMAX and TAMEAN (Manual or Auto Trace)
Two Velocities with the Time Interval and Acceleration
between them
• Time Interval
• Volume Flow
NOTE: The following instructions assume that you do the following:
1. In the B-Mode part of the display, scan the anatomy you
want to measure.
2. Go to the Doppler Mode part of the display.
3. Press Freeze.
Velocity
To measure velocity:
1. Press Measure; an active caliper with a vertical dotted line
displays.
2. To position the caliper at the desired measurement point,
move the Trackball.
3. To complete the measurement, press Set. The system
displays the velocity measurement in the Results Window.