Intensity (Echo level) measurement
To make an echo level measurement:
1. Press Measure.
2. Press the top Trackball key to select Intensity. A caliper
displays.
3. To position the caliper at the start point, move the Trackball.
4. To fix the trace start point, press Set. The caliper changes to
an active caliper.
5. To trace the measurement area, move the Trackball around
the anatomy. A dotted line shows the traced area.
6. To complete the measurement, press Set. The system
displays the echo level, as EL __ dB, in the Results Window.
NOTE: The echo level measurement is only available on a frozen
image, not on a B-paused image.
NOTE: Echo Level is not available through the factory default. To
enable echo level, modify the Measure Key Sequence preset,
found in the Utility -> Measure -> Advanced preset.