Circumference and area (spline trace) measurement (continued)
Edit the spline trace
1. Select Cursor Select. The spline trace changes to green
and all points appear on the trace as yellow.
A pick-caliper appears on the center of the image and the
message “Edit spline trace” displays at the bottom of the
screen.
NOTE: The pick-caliper is used to select and move the trace points.
Figure 2-3. Edit spline trace
Select Cursor Select again. The trace is deactivated
(changes to yellow) and all points, including the pick-caliper,
are removed.
If the previous/next fixed caliper exists on the image, it is
activated.
NOTE: Pressing Clear at this time removes all points and the trace
graphic.
2. Move the pick-caliper to the desired point and press Set.
The point is activated and turns green.
3. Move the point to the desired position and press Set. The
point is fixed and turns yellow. The pick-caliper appears on
the center of the image.
NOTE: The spline trace is updated at run time.
NOTE: To remove a point, press Clear while moving the point. The
trace turns green and the remaining points continue to be
shown as yellow. If there are less than three points, the
spline trace is removed.
4. Press Set again. All points are removed from the trace and
the trace is shown as yellow.

Table of Contents

Other manuals for GE LOGIQ E9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E9 and is the answer not in the manual?

GE LOGIQ E9 Specifications

General IconGeneral
TypeUltrasound System
3D/4D ImagingYes
Transducer Ports4 active ports
ConnectivityDICOM, USB, Ethernet
ApplicationsAbdominal, Vascular, Musculoskeletal, Small Parts, Urology, Pediatric, Cardiac, Obstetrics/Gynecology
TransducersLinear, Convex, Phased Array, Volume
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
ElastographyYes
Advanced FeaturesB-Flow, Volume Navigation
PortabilityCart-based system
Display19-inch high-resolution LCD monitor

Related product manuals