Measurement and Analysis
LOGIQ E9
User Guide
2-23
Direction 5454884-100 English
Rev. 1
Circumference and area (spline trace) measurement
To trace the circumference of a portion of the anatomy and
calculate its area:
1. Press Measure.
2. Press the top Trackball key to select Spline Trace; a caliper
displays.
3. To position the first caliper at the start point, move the
Trackball.
4. To fix the trace start point, press Set. The first caliper turns
yellow. The second caliper appears at the same position as
the first caliper and is green.
NOTE: When pressing the Clear key once, the second caliper
disappears and the first caliper is activated.
If Clear is pressed again, the first caliper disappears and the
Spline trace is cancelled.
5. To position the second caliper, move the Trackball and
press Set. The third caliper appears at the same position.
NOTE: The Clear key functionality is the same as noted in the
previous step.
The spline trace requires at least three points to draw the
trace. Continue setting the points of the trace until the
desired points are set.
6. Press Set again after the last caliper is fixed to finalize the
spline trace. All points are removed from the line and the
spline trace turns yellow.
NOTE: Pressing Set twice finishes the trace measurement.
If Clear is pressed twice when more than 3 points exist on
the trace, all points are removed and the first caliper again
displays.
NOTE: Spline trace is not available through the factory default. The
system defaults to trace. To enable spline trace, modify the
Measure Key Sequence preset found in Utility -> Measure ->
Advanced preset menu.

Table of Contents

Other manuals for GE LOGIQ E9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E9 and is the answer not in the manual?

GE LOGIQ E9 Specifications

General IconGeneral
TypeUltrasound System
3D/4D ImagingYes
Transducer Ports4 active ports
ConnectivityDICOM, USB, Ethernet
ApplicationsAbdominal, Vascular, Musculoskeletal, Small Parts, Urology, Pediatric, Cardiac, Obstetrics/Gynecology
TransducersLinear, Convex, Phased Array, Volume
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
ElastographyYes
Advanced FeaturesB-Flow, Volume Navigation
PortabilityCart-based system
Display19-inch high-resolution LCD monitor

Related product manuals