EasyManua.ls Logo

GE LOGIQ E9

GE LOGIQ E9
256 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Safety Precautions
LOGIQ E9
User Guide
4-9
Direction 5454884-100 English
Rev. 1
Mechanical hazards
The use of damaged probes or improper use and manipulation
of intracavity probes can result in injury or increased risk of
infection. Inspect probes often for sharp, pointed, or rough
surface damage that could cause injury or tear protective
barriers.
The use of damaged probes can result in injury or increased risk
of infection. Inspect probes often for sharp, pointed, or rough
surface damage that could cause injury or tear protective
barriers.
Electrical
Hazard
A damaged probe can also increase the risk of electric shock if
conductive solutions come in contact with internal live parts.
Inspect probes often for cracks or openings in the housing and
holes in and around the acoustic lens or other damage that
could allow liquid entry. Become familiar with the probe's use
and care precautions outlined in Probes and Biopsy.
CAUTION
Ultrasound transducers are sensitive instruments which can
easily be damaged by rough handling. Take extra care not to
drop transducers and avoid contact with sharp or abrasive
surfaces. A damaged housing, lens or cable can result in
patient injury or serious impairment or operation.
CAUTION
Do not use with Defibrillator.
This equipment does not have a defibrillator approved applied
part.

Table of Contents

Other manuals for GE LOGIQ E9

Related product manuals