Probes
LOGIQ E9
User Guide
1-57
Direction 5454884-100 English
Rev. 1
Endocavitary Probe Biopsy Guide Assembly - Representative Example
Preparation
To prepare the endocavitary probe for use:
1. Remove the probe from the box and carefully examine it for
any damage.
2. If the biopsy guide is to be attached, use the filling removal
tool to clean out the attachment area on the probe head.
Figure 1-38. Attachment Filling Removal
a. Probe Head
b. Attachment
c. Filling Removal Tool
3. Clean, then disinfect the probe.
NOTE: Ensure that protective gloves are worn.
WARNING
DO NOT use the needle with the catheter (soft tube). There is a
possibility of breaking the catheter in the body.
CAUTION
Before inserting the needle, scan the patient to determine the
correct puncture depth and site. Only the sterile/sanitary
sheath and rubber band are on the probe during the pre-needle
placement scanning.
a
b
c