EasyManuals Logo

GE LOGIQ E9 User Guide

GE LOGIQ E9
256 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #86 background imageLoading...
Page #86 background image
Getting Started
1-74
LOGIQ E9
User Guide
Direction 5454884-100 English
Rev. 1
Patient Screen (continued)
Maintaining the Patient Database
1. Before starting EZBackup, select “Unlock All” on the
Utility -> Admin -> Logon screen.
Prepare the unformatted CD-R or DVD-R before EZBackup.
NOTE: Pre-formatted CD-R or DVD-R cannot be used for
EZBackup.
2. First perform EZBackup and then Backup (Patient Archive
and Report Archive).
3. Go to the patient screen, select the patients/exams to
delete. Select “Delete” to delete the selected data.
NOTE: Removing image data with the “EZMove” function does not
reduce the patient number in the database.
NOTE: Ensure that all patients are exported or backed up BEFORE
deleting them.
We recommend attaching the patient list to the EZBackup
media. Insert the media and select DICOM CD Read for the
dataflow (if you use a USB drive, select DICOM USB Drive
Read). Press the right Set key to select the patient. Press the
left Set key to bring up the print menu. Select Print and press
the right Set key to print the patient list.
CAUTION
To maintain optimum system performance and to safeguard
patient data, keep the total number of patients in the database
below 1,000.
To reduce the total number of patients in the database, perform
the following procedure.

Table of Contents

Other manuals for GE LOGIQ E9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E9 and is the answer not in the manual?

GE LOGIQ E9 Specifications

General IconGeneral
TypeUltrasound System
3D/4D ImagingYes
Transducer Ports4 active ports
ConnectivityDICOM, USB, Ethernet
ApplicationsAbdominal, Vascular, Musculoskeletal, Small Parts, Urology, Pediatric, Cardiac, Obstetrics/Gynecology
TransducersLinear, Convex, Phased Array, Volume
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
ElastographyYes
Advanced FeaturesB-Flow, Volume Navigation
PortabilityCart-based system
Display19-inch high-resolution LCD monitor