EasyManua.ls Logo

GE LOGIQ P6

GE LOGIQ P6
478 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 5245279, REVISION 3 LOGIQ™ P6/P6 PRO SERVICE MANUAL
8 - 100 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-47 Read and Write Serial Info
8-2-47-1 How to start serial no read/write program
1.) Plug the service dongle on the usb port on rear panel.
2.) Turn on the system.
3.) When the Start Application dialog box appears, press [F2] before it automatically starts
Echoloader.)
NOTE: If maintenance access box does not appear and the scanning mode starts, just insert the
Service Dongle and press “CTRL+ALT+R”.
4.) When the Maintenance Access dialog box pops up, type in the service password.
Figure 8-147 Start Appllication Dialog Box
Figure 8-148 Maintenance Access Dialog Box

Table of Contents

Other manuals for GE LOGIQ P6

Related product manuals