EasyManua.ls Logo

GE LOGIQ P6

GE LOGIQ P6
478 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 5245279, REVISION 3 LOGIQ™ P6/P6 PRO SERVICE MANUAL
8 - 114 Section 8-4 - Software Installation
8-4-1 Installing Base Image Software (cont’d)
8.) Press Enter key and then the following screen will be shown. (If you want to abort installation
procedures, press CTRL+C.)
9.) Press A key to install the complete disk.
Figure 8-169 Base Image Software Installation Menu
NOTICE
Select A when installing the new HDD .
Select B when replacing C:\ partition only. It means all customer data are NOT deleted. The Ghost pro-
cedure is performed with data of the drivers D, E, F and Z left."
Select C when replacing C:\ partition and Z:\ partition. It means all customer data are NOT deleted. The
Ghost procedure is performed with data of the drivers D, E and F left. Z:\ partition used for reloading
C:\ partition and Application software will be replaced.
Drive Partition Usage
(GRUB) grub data
C System Files and Application
D User Defined and Preset
E Patient Archive data
FSwap
Z C:\partition Repository for Reloading Function
Table 8-48 Drive Partition

Table of Contents