DIRECTION 5245279, REVISION 3 LOGIQ™ P6/P6 PRO SERVICE MANUAL
Section 8-4 - Software Installation 8 - 133
8-4-4-5 Setting the VCR (if necessary)
1.) Select Utility > System>Peripheral tab
2.) Under Video Settings, PAL or NTSC.
3.) Under VCR, select Composite or S-Video.
4.) Under VCR, select Panasonic LQ-MD800 or Mitsubishi MD3000, as appropriate.
5.) Click on Save.
6.) Shut down the scanner, then restart the scanner.
7.) Verify that the VCR can be properly operated..
8.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Figure 8-198 Utility > System>Peripheral
NOTICE
Be sure to connect the converter and cablebefore booting up the system. Otherwise the LOGIQ™ P6/
P6 Pro system does not recognize the peripheral devices
Table 8-52 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test Debrief Script
Section 4-3-1
Power On/Boot Up
“Service Manual, Direction
5245279, Rev 3+, Section 8-4-4. Equipment
passed all required tests and is ready for use. “
Section 4-3-2
Power Off / Shutdown
Section 4-5
Peripheral Checks