GE HEALTHCARE
DIRECTION 5245279, REVISION 3 LOGIQ™ P6/P6 PRO SERVICE MANUAL
Section 8-4 - Software Installation 8 - 145
8-6-2-1 Perform Software Reloading in the Service Platform (cont’d)
3.) Press “Time Zone”
4.) Register the current Time Zone
5.) Press “Cancel”.
6.) Log onto the Service Platform.
7.) Go to “Utilites”->”Common Utilites”.
8.) Go to "Utility"->"System"->"About"
Verify that Software version and software build version are same with the ones before the Software
Reloading.
Figure 8-217 Time Zone
Current Time Zone
Table 8-1 Current Time Zone