Preparing the Recording
48 SEER™ 1000 2067634-077 Revision B
each time wireless communication takes place,
if the cable status changes (e.g., cable attached or removed), or
if the Event Button is pressed but no recording starts because of low
battery or open lead (error sound).
NOTE
Once the recording has automatically started, it is not possible to ac-
cess the recorder demographic and ECG data using an app until the
recording has been stopped and the data deleted via the MARS or
CardioDay software. However, the recorder can be reset using the PC
App as explained in section “Resetting the Recorder with the Mi-
crosoft Windows App” on page 90.
The Apple iOS version of the application does not allow the recorder
to be reset.
NOTE
While the auto-start function helps to prevent the patient from carry-
ing the recorder unnecessarily, we recommend that you manually
start each recording. The automatic start function should only be
considered as a backup feature.
Auto-Switch-off
If the recording has not been started, the recorder will automatically
switch off after eight minutes if no ECG is detected and if no communica-
tion (wireless or via USB) takes place.
The eight-minute timer is retriggered
each time wireless communication takes place,
if the cable status changes (e.g., cable attached or removed), or
if the Event Button is pressed but no recording starts because of low
battery or open lead (error sound).