EasyManua.ls Logo

GE SEER 1000 - Troubleshooting; Recorder Troubleshooting

GE SEER 1000
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8
2067634-077 Revision B SEER™ 1000 83
8 Troubleshooting
This section explains error signs and troubleshooting recommendations.
Recorder Troubleshooting
Trouble Indicators
Cause
Recommendation
The Event Button LED
flashes yellow once
every second.
An error tone sounds
when you try to start the
recording.
Battery power is low.
The battery is not fully charged
or the wrong battery type has
been inserted. Replace with
new alkaline battery or NiMH
rechargeable battery.
The Hookup Guidance
LED flashes yellow once
every second.
An error tone sounds
when you try to start the
recording.
Electrode or ECG cable is not
connected.
Connect the ECG cable or re-
connect loose electrodes.

Table of Contents

Related product manuals