Troubleshooting
2067634-077 Revision B SEER™ 1000 85
Application Troubleshooting
Message
Recommendation
“A technician with the same identifier exists al-
ready.”
Enter a new technician identifier.
“An error occurred while connecting. You
should try again.”
Try again to connect to the recorder.
“Battery level too low to start a recording.”
For the recording duration selected, the
battery is not sufficiently charged or the
wrong battery type has been inserted.
Insert a new alkaline or a fully charged
NiMH battery, or reduce the recording
duration.
“Battery low. Recorder is shutting down.
Insert a new alkaline or a fully charged
NiMH battery.
“Data could not be deleted. Please check if the
SEER recorder is connected to another soft-
ware.
If the recorder is currently connected to
an evaluation software via USB, e.g., to
transfer a recording, wait until this data
exchange is terminated and then discon-
nect the USB cable before you re-try to
delete the data.
Data stored on recorder. Do you want to select
a different recorder?
Delete the data stored on the recorder or
download it to the an evaluation soft-
ware.
Alternatively, select another recorder.
“Deleting recording failed.”
Try again as the wireless communication
may have been interrupted.