Troubleshooting
92 SEER™ 1000 2067634-077 Revision B
Failure number
(Count of LED
flashes)
Failure
1
The firmware CRC check fails. The board needs to be reprogrammed.
2
The internal RAM is defective or has a soldering problem.
3
One or more of the different internal clocks fails.
4
The communication between the µC and the NAND-Flash fails.
5
The NAND-Flash has too many bad blocks.
6
The communication to the Bluetooth module fails.
7
The RTC (Real Time Clock) is in an error state. Restart of the device may fix
the problem.