Introduction
1–6
MP-129F Water-Saver Pressure values are evaluated by the sterilizer software to determine
whether the water-saver or the standard ejector in the sterilizer provides the
vacuum. The water-saver configuration dipswitch (DS14 on the sterilizer’s
DISPSWITCHES menu) must be set to 1 in order for the water-saver to
operate. While the pressure in the chamber is above -5 psig for vacuum
cycles, or above atmospheric pressure for gravity cycles, the sterilizer
software selects the standard ejector. When below these values, the
software turns on the water saver.
Figure 1–4. MP-129F WATER-SAVER SYSTEM
Ejector
Pump
1SV
Manual
Drain
Valve
Copper
Tubing
1S
Exhau
from Sterilizer
Check
Valve
Cold Water
Temperature
Switch
Customer Pipin
El
tri
a
115 VAC
350545-AE
I/F Box
11A
Water-Saver ON
Output DO14
from Sterilzer
Water-Saver ON
15SV
Click Here To Open Parts List & Wiring Rev F