Maintenance and Troubleshooting
4ā2
CLEANING AND SANITIZING THE TANK
To clean and sanitize the water-saver tank:
1. Disable the water-saver:
a. Turn OFF the bypass switch S1.
b. Disconnect the high-voltage power supply to the pump.
On MP-129F units, disable the water-saver by setting configuration switch
DS14 to 0. Refer to the sterilizer service manual for procedure.
2. Remove the large plug in the top of the tank.
3. Add sanitizing agent to the tank water.
Use an off-the-shelf, non-toxic, non-caustic water tank
sanitizer and cleaner.
4. Agitate the water solution. Use a long-handled brush if
necessary for proper cleaning.
5. Drain the tank.
6. Enable the water-saver:
a. Connect the high-voltage power supply to the pump.
b. Turn ON the bypass switch S1.
On MP-129F units, enable the water-saver by setting configuration switch
DS14 to 1.
7. Verify that the tank refills.
8. Replace the large plug.
NOTE
NOTE