350545 2–1
Section 2 MP-129 Installation
Follow the instructions in this user manual when installing an MP-129
Water-Saver to any of the following sterilizers:
• 100 and M/C 3500 Series (MP-129A)
• 200 and M/C 3600 Series (MP-129B)
• M/C 4233 (42") (MP-129C)
• M/C 4233 (76") (MP-129D)
If you are installing a MP-129F Water-Saver to a 400/500, 700
or 800 Series sterilizer, follow the instructions in Section 3.
High-vacuum sterilizers (x33 or xx33 models) benefit most from the use of
the water-saver. However, the water-saver can be installed on any Getinge
USA sterilizer that uses a water ejector as its vacuum source.
Bills of Materials (BOMs) and drawings necessary for assembling this kit are
included with the parts.
• Use the BOMs as a checklist to verify receipt of all parts.
• Use the BOMs and the drawings to assemble the parts per these
instructions.
• Retain all BOMs and drawings with the unit for future reference
CAUTION
NOTES