MP-129F Installation
3ā4
LOCATING, MOUNTING, AND ANCHORING
Location Refer to installation drawing HS3472 (included in water-saver package).
The water-saver can be placed on the floor or mounted to the ceiling in any
convenient location adjacent to the sterilizer. A 20-foot length of copper tub-
ing, with connection fittings is supplied for connecting the water-saver to the
sterilizer.
Locating the water-saver within the connected reach of the copper tubing
limits the positioning to approximately 15 feet from the sterilizer. If the water-
saver is to be located on the side of the sterilizer opposite the connection
piping, the distance may be somewhat reduced.
When mounting to an overhead structure, positioning the vacuum port of the
water ejector on the water-saver 10 feet above the floor is about the maxi-
mum height that permits the 20-foot copper tubing to reach the sterilizer.
Mounting and Anchoring Mounting brackets are provided to secure the unit to a wall or to mount it to
the ceiling. Mounting holes in the bottom plate are provided for securing the
unit to the floor.
Structures for ceiling-mounted units must be able to support a minimum of
250 lbs. (113 Kg) (150 lbs. [68 Kg] plus water, hardware, and safety margin).
Use 3/8" diameter bolts with appropriate anchoring devices to secure the
water-saver.
For seismic anchoring requirements, worst case center of gravity location,
and weight, refer to drawing No. 531199 (located in Section 5).
The water-saver can stand stationary of its own weight if anchoring is not
required or in areas where accidental tipping is not a possibility.
All mounting hardware other than the mounting brackets provided on the
water-saver is the responsibility of the customer installer.