A short-circuit has been detected
at the test socket between L and
N.
Determine whether or not the de-
vice under test is defective.
In the case of DUTs which are in-
tended for operation at an outlet
that’s protected with a 16 A fuse, a
short-circuit may be detected
under certain circumstances if, for
example, they include a PTC resis-
tor (e.g. large floodlights). Be sure
to use a 3-phase test adapter in
order to test devices of this sort
(e.g. the AT3-IIIE).
You can deactivate this short-cir-
cuit test in the sequence parame-
ters at your own risk.
A short-circuit has been detected
at the test socket between L/N and
PE.
Determine whether or not the de-
vice under test is defective. Repeat
the visual inspection.
General Application Errors
The inspector to be deleted is cur-
rently selected and thus cannot be
deleted!
Activate a different inspector before
deleting.
The 25 A measurements takes too
long.
or
The 25 A measurement has been
executed too often (without
pauses).
Wait until the test instrument has
cooled down and then restart the
measurement.
Error Messages Possible Causes Corrective Measures

Table of Contents

Other manuals for Gossen MetraWatt SECUTEST PRO

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gossen MetraWatt SECUTEST PRO and is the answer not in the manual?

Gossen MetraWatt SECUTEST PRO Specifications

General IconGeneral
BrandGossen MetraWatt
ModelSECUTEST PRO
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals