GMC-I Messtechnik GmbH 75
Database Processing Error
– One of the fields was filled in with
invalid content while processing an
existing database object.
➭ Please be certain to complete all
mandatory fields (identified in red).
➭ If necessary, check your entries to
the fields for invalid special charac-
ters.
– The ID field was not filled in while
creating a new device.
➭ Fill in the ID field.
– There’s already an object with the
same ID under the “Customer”
database object.
An incorrect barcode has been
selected.
➭ Assign another ID.
Error while writing the “.secu” file to the
USB flash drive
There’s not (no longer) enough avail-
able memory space on the storage
medium.
– In particular in the case of FAT16
formatted USB flash drives: Too
many files on the USB flash drive
– Power consumption of the utilized
USB flash drive exceeds 500 mA.
– The USB flash drive has been dis-
connected during data import.
– The USB flash drive is defective or
incompatible with the SECUTEST.
Make sure that at least 100 MB is
available on the USB flash drive or
delete any data files which are no lon-
ger required.
➭
If the problem persists, save the data
from the USB flash drive on another
storage device and reformat the USB
flash drive (FAT32).
➭ Only use USB flash drives with
power consumption of less than
500 mA in combination with the
SECUTEST.
➭ Make sure that the USB flash drive
is not disconnected or moved until
the entire data export process has
been completed.
➭ If none of these measures results in
improvement, replace the USB
flash drive. A list of tested USB
flash drives is included in section
14.3.
Error Messages Possible Causes Corrective Measures