EasyManuals Logo

GW Instek GPP-3060 User Manual

GW Instek GPP-3060
218 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #131 background imageLoading...
Page #131 background image
REMOTE CONTROL
131
Description
< function >: VOLTage,CURRent,POWEr
Example
:MEASure:VOLTage:ALL?
Reads back the all channels voltage value.
Note
About CH3 , The query voltage will return the set
value, current and power will return zero.
Command
VOUT<X>?
Function
Returns the actual output voltage.
Description
X 1= CH1, 2= CH2
Example
VOUT1?
Returns the CH1 output voltage.
Note
About CH3, The query voltage will return the set
value.
Command
IOUT<X>?
Function
Returns the actual output current.
Description
X 1=CH1, 2=CH2
Example
IOUT1?
Returns the CH1 output current.
Note
About CH3 , The query current will return zero.