GPP-3060/6030 User Manual
98
Firmware Upgrading
When to
Upgrade
Firmware
When system is failure, request by customer GW
Instek. When the system fails, firmware can
requested by GW Instek customers.
Upgrade
Requirement
Firmware file
Supplied by GW Instek
Operation
After plugging in flash drive, the corresponding
symbol appears in the status bar.
Press System button;
Press F3 (Setting) button;
Press F4 (Upgrade) button;
Rotate the scroll wheel to select corresponding
file;
Press the F4 (Recall) button and start to upgrade;
The procedure is done when the message,
upgrading successful, shows up.
Note
Flash drive only support USB2.0/USB3.0,FAT32
file system,16G MAX.