APPENDIX
215
Note
Tracking is not supported in LOAD mode.
Resolution
Voltage
Current
programming 1mV ,readback 0.1mV (GPP-3060)
programming 2mV ,readback 0.1mV (GPP-6030)
programming 0.2mA,readback 0.1mA) (GPP-3060)
programming 0.1mA,readback 0.1mA) (GPP-6030)
Accuracy
Ammeter
full scale:6.2A(GPP-3060),3.2A (GPP-6030)
programming 5 digits readback 5 digits
Voltmeter
full scale:32V(GPP-3060), 62V(GPP-6030)
programming 5 digits,readback 6 digits
Setting
accuracy
Voltage: ± (0.03% of reading + 10mV)
Current: ± (0.3% of reading + 10mA)
Readback
accuracy
Voltage: ± (0.03% of reading + 10mV)
Current: ± (0.3% of reading + 10mA)
Bindpost
Output
voltage
1.8V/2.5V/3.3V/5.0V, ±5%
Output
current
5A
CH3
Line
regulation
3mV
Load
regulation
5mV
Ripple &
noise
2mVrms (5Hz - 1MHz)
Transient
recovery
time
100µs
(50% load changeminimum load 0.5A)
USB Port
Output
1.8V/2.5V/3.3V/5.0V, ±0.35V, 3A
Warning
The output current from the 2 terminals should Not
exceed 5A.
Load Mode
Display
Voltage
1-32.00V (GPP-3060)
1-62.00V (GPP-6030)
Current
0-6.200A (GPP-3060)
0-3.200A (GPP-6030)
Power
0-50.00W(CH1/CH2)
CV Mode
CH1/CH2
1.500V - 32.00V (GPP-3060)
1.500V - 62.00V (GPP-6030)