Mechanical installation
24 A0785-150 HBM: public T10FS
5.4 Mounting the rotor
Additional installation notes for the speed measuring
system can be found in Chapter 5.7, Page 32.
Notice
Usually the rotor identification plate is no longer visible
after installation. This is why we include with the rotor
additional stickers with the important ratings, which you
can attach to the stator or any other relevant test‐bench
components. You can then refer to them whenever there
is anything you wish to know, such as the calibration
signal. To explicitly assign the data, the identification
number and the measuring range are engraved on the
rotor where they can be seen from outside (see Fig. 5.3).
1. Prior to installation, clean the plane surfaces of the
measurement flanges and counter flanges. For safe
torque transfer, the surfaces must be clean and free
from grease. Use a piece of cloth or paper soaked in
solvent. When cleaning, make sure that you do not
damage the transmitter coils.