EasyManuals Logo

HBM T10FS User Manual

HBM T10FS
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #94 background imageLoading...
Page #94 background image
Specifications
94 A0785-150 HBM: public T10FS
Nominal (rated) torque M
nom
10 k5 k3 k2 k1 k500200100Nm
Rel. standard deviation of
reproducibility
according to DIN 1319,
relative to the variation of
the output signal
Frequency output % <
±0.03
<±0.02
Voltage output % <±0.03
Calibration signal approx. 50% of M
nom
; more precise value on
identification plate
Tolerance of the shunt
signal, related to M
nom
% <±0.05
1)
RS-422 complementary signals; factory setting version SF1/SU2
2)
Factory setting version KF1 (changeover not possible)
3)
Output signal range in which there is a repeatable correlation between torque and output signal.