32
pretreatment cartridge this chapter to operate.
Purification cartridge is installed in the left chamber of the main unit,
Remove exhausted purification cartridge.
Take out the new ultra-purification cartridge, and make sure the order No. is
CR-SP302M.
Install the purification cartridge into the main unit, Install the side door.
Ultra-purification cartridge replacement(CR-SP303M)
Open the front door of the Smart-Mini system. Please refer to 4.3.5of chapter 4.
Open the front door, then you can see upper and lower latches inside the main unit. Take
out of the latches. Please refer to the figure A in section 4.3.5 of chapter 4.
Remove the exhausted ultra-purification cartridge from the main unit.
Get the new purification cartridge, and make sure the order number is CR-SP303M.
Install the new purification cartridge into the main unit. Please refer to the section 4.3.5
of chapter 4.
Install the front door.
GO back to the control panel, press the "Stop /
" keypad to confirm the replacement
process is done.
After finishing the purification cartridges replacement, you can press "Disp. / ▲ " or
"Flush /
▼
" button the selection other components replacement, or press "Menu" button to
return to the upper level of menu. or you can continuously press "Menu" keypad to exit the
maintenance setting, the system will reboot.
Newly installed ultra-pure purification cartridges and cartridges need to be washed for 2
to 4hours to reach the best condition. When the new cartridge is installed, the ultra-pure water
resistivity will gradually be increased. Once you reach the setting value (manufacture setting
is 16.0MΩ.cm), the ultra-pure water can be produced. Then you can press the "Stop /
"
keypad to stop the resin flushing cycle and the system is into the standby mode.
When the new cartridge is in the flushing cycle, Place a water container under the
ultra pure water outlets, you should be to view the system LCD displays. When the
quality of water is going to meet the pre-setting value hold the container in right
position to avoid damage to other equipments.