EasyManua.ls Logo

Hettich EBA 280 S - Page 164

Hettich EBA 280 S
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DE
EN
164/165
1139 1054-A
---
0553 ---
0518
Winkelrotor 12-fach /
Angle rotor 12-times
35°
3)
0701
Kapazität / capacity ml
5 12 1,6 – 5,0 1,1 – 1,4 2,6 – 3,4 2,7 - 3 4 15
Maße / dimensions x L
mm
12 x
75
13 x
75
17 x 102
13 x 75 8 x 66 13 x 65 11 x 66 12 x 60 17 x 100
Anzahl p. Rotor / number p. rotor
12
Drehzahl / speed RPM
6000
RZB / RCF
2)
3300 4146 3300 3300 3300 3300 3260 4146
Radius / radius mm
82 103 82 82 82 82 81 103
9 (97%)
sec
16
9
sec
16
Probenerwärmung/Sample
temp. rise
K
1)
6
1139
1058 1063 6305
0509 0507
Winkelrotor 12-fach /
Angle rotor 12-times
35°
Kapazität / capacity ml
4,5 -
5
4,9
7,5 – 8,5
9 -
10
10
4 - 7
8
8,5 -
10
0,5 15 4
Maße / dimensions x L
mm
11 x
92
13 x
90
15 x 92
16 x
92
15 x
102
13 x 100
16 x
125
16 x
100
10,7 x 36
17
x
120
17
x
100
10 x 88
Anzahl p. Rotor / number p. rotor
12 12 12 12 6 12 12 6 12 12
Drehzahl / speed RPM
6000
RZB / RCF
2)
4146 4146 4146 4146 4146 2777 4146 3502
Radius / radius mm
103 103 103 103 103 69 103 87
9 (97%)
sec
16
9
sec
16
Probenerwärmung/Sample
temp. rise
K
1)
6
1) Probenerwärmung bei maximaler Drehzahl und
1 Stunde Laufzeit
1) Sample temp. rise during maximum speed and 1 hour
running time
2) Zentrifugiergefäße aus Glas nur belastbar bis RZB 4000 2) Centrifuge containers of glass will not stand RCF values
exceeding 4000
3) ohne Deckel 3) without lid

Table of Contents

Related product manuals