24.3 Autoclaving
The following accessory can be autoclaved at 121°C / 250°F (20 min):
 Swing-out rotors without hangers
 Reductions
No statement can be made about the degree of sterility.
Autoclaving accelerates the ageing process of plastics. In addition, autoclaving may discolour plastics.
24.4 Centrifuge containers

With leakiness or after the breakage of centrifuging containers broken container parts, glass splinters and leaked
centrifugation material are to be completely removed.
 The rubber inserts as well as the plastic sleeves of the rotors are to be replaced after a glass breakage.
Remaining glass splinters cause further glass breakage!
 If this concerns infectious material, a disinfection process is to be executed immediately.